  technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 1 15 0 a as 24 f/ b highlights & features C meet doe level v i and erp lot 7 C safety approvals to iec 60601 - 1 3 .1 rd ed. & iec 60950 - 1 C compliant with iec 60601 - 1 - 2 4th ed. requirements C low touch current (<0.1ma normal & ,0.3ma single fault) C over - voltage/load/temper ature & short circuit protections C 1 million hours mtbf C 2 mopp isolation C 3 years warranty safety standards cb certified for worldwide use model number: mds - 15 0 a a s 24 f/ b unit weight : 1.1 kg ( 39.6 o unces ) dimensions ( w x l x h ) : 8 5 .0x1 7 0 .0x 40 . 0 mm 3.3 x 6.7 x1. 6 in general description the mds series of external p ower s upply come with universal ac input at 90 vac to 264vac. other features include low touch current , risk management report available and the electric shock protection comply with 2 x mopp . the mds series is certified for emc standards according to en 55011 for industrial, scientific and medical (ism) radio - frequency equipment and en 55022 for industrial technology equipment (ite) radio - frequency equipment . in addition, only recognized japa nese capacitors are used. the mds series come with both medical and ite s afety approvals including ul/ cul /c c c/ce and cb certification and are fully compliant with rohs directive 2011/65/eu for environmental protection. model information medical ac - dc adap ter model number input v oltage range output v oltage output cur r ent mds - 15 0 a as 24 f/ b 90 - 2 64 v ac 24 vdc 6.25 a model numbering 15 0 max wattage in the p roduct s eries. maybe lower at some voltage. 15 0 1 5 0 w mds delta medical power supply - a as family code 24 output v oltage single output: 24 for 24 v f/b b
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 2 specifications input ratings / characteristics nominal input v oltage 100 - 2 40 v ac input v oltage range 90 - 2 64 v a c nominal input f r equency 50 - 60hz input f r equency range 47 - 63hz input cur r ent (max.) 4 a @ 1 15 v ac, 2 a @ 2 3 0 v ac e f ficiency (typ.) 90 . 7 % , reference fig.1 average efficiency (min) 88% @ 115vac and 87% @ 230vac standby power (max.) 0.21w @ 115vac and 0.5 w @ 230vac inrush cur r ent (typ.) 50 a @ 1 15 v ac, 100 a @ 2 3 0 v ac touch cur r ent (max.) 0.1ma @ 264 v ac nc 1) , 0. 3 ma @ 264 v ac sfc 2) 1) n c: normal condition 2) sfc: single fault condition fig.1 efficiency versus output load output ratings / character istics nominal output v oltage 24 vdc total regulation 5 % output cur r ent 6.25 a output power 1 5 0 w line regulation (max)
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 3 (a) 115v (b) 230v fig. 2 ripple & noise example , 20mhz bw ripple & noise measurement circuit mechanical case chassis pc case co ver pc dimension s (w x lx h) 8 5 .0 x 1 7 0.0 x 40 . 0 mm unit weight 1.1 k g ( 39.6 o unces ) indicator na cooling system na terminal input c14 socket output 4pin din / barrel type / tuning fork type environment surrounding air temperature operating 0c t o + 4 0c storage - 40 c to + 8 5c operating humidity 1 0 - 9 5 % rh (non - condensing) operating altitude 3 ,000 m eters (9842.52 feet ) shock test (non - operating) 50g, 11ms, 3 shocks for each direction vibration (operating) 5 - 500hz, 2.09grms, 20 minute for e ach three axis 0 . 4 7 u f c e r a m i c 1 0 u f e l e c t r o l y t i c p s u l o a d t e s t f i x t u r e 5 c m s c o p e
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 4 protections overvoltage (max ) 150% , latch mode over load / over current (max ) 1 3 0 % of rated load current, hiccup mode, ( non - latching , auto - recovery) over temperature hiccup mode, ( non - latching , auto - recovery) short circuit hiccup mod e, ( non - latching , auto - recovery) protection against shock class i for f series class i i with functional earth for b series reliability data mtbf (typ.) 1 million hours based on telecordia sr - 332 safety standards / directives medical safety ie c60601 - 1 3 rd and 3.1 rd ed ition cb report iec60601 - 1 e dition 3.1 rd (2012), en60601 - 1 (2006) + a11 + a1 + a12, can/csa - c22.2 no. 60601 - 1:14, ansi/aami es60601 - 1:2005/(r)2012 ite safety iec60950 - 1 (ed.2,2005), gb4943.1 - 2011, gb9254 - 2008, gb17625.1 - 2003 ce mdd directive 93/42/eec material and parts rohs directive 2011/65/eu compliant galvanic isolation input to output 4000 vac input to ground 1500 vac output to ground 500 vac
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 5 emc ( compliant with iec 60601 - 1 - 2 4th ed. requirements ) emc / emission s en550 11/ en550 22, fcc title 47 : class b harmonic current emissions iec61000 - 3 - 2 meet class a limit immunity to voltage flicker iec 61000 - 3 - 3 electrostatic discharge iec61000 - 4 - 2 level 4 criteria a 1) air discharge: 15 kv contact discharge : 8 kv radiated field iec61000 - 4 - 3 criteria b 2 ) 80mhz - 2700mhz, 10v/ m am modulation 385mhz - 5785mhz, 28v/ m pulse mode and other modulation electrical fast transient / burst iec61000 - 4 - 4 level 3 criteria a 1) : 2kv surge iec61000 - 4 - 5 level 3 criteria a 1) com mon mode 3 ) : 2kv differential mode 4 ) : 1kv conducted iec61000 - 4 - 6 level 2 criteria a 1) 150khz - 80mhz, 3 vrms , 6vrms at ism bands and amateur radio bands power frequency magnetic fields iec61000 - 4 - 8 criteria a 1) magnetic field strength 3 0 a/ m voltage dip s iec61000 - 4 - 11 criteria a 1) 0% u t , 0.5 cycle ( 10ms ) , 0 o /45 o /90 o /135 o /180 o /225 o /270 o /315 o /360 o criteria a 1 ) 0% u t , 1 cycle ( 20ms ), 0 o criteria a 1 ) 70 % u t , 25 cycle ( 500ms ), 0 o criteria b 2) 0 % u t , 250 cycle ( 50 0 0ms ), 0 o 1) criteria a: normal performan ce within the specification limits 2) criteria b : output out of regulation, or shuts down during test. automatically restore to normal operation after test. 3) asymmetrical: common mode (line to earth) 4 ) symmetrical: differential mode (line to line)
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 6 item device description 1 cover 2 chassis 3 power cord 4 socket block diagram dimensions w x l x h : 85.0 x 170 .0 x 40.0 mm mm
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 7 functions start - up time the time required for the output voltage (vo) to reach 90% of its set value, after the input ac voltage is applied. rise time the time required for the output voltage (vo) to change from 10% to 90% of its steady state value. hold - up time hold up time is the time when the ac input collapses and output voltage retains regulation for a certain period of time. the time required for the output to reach 95% of its s et value, after the input voltage is removed. graph illustrating the start - up time, rise time, and hold - up time
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 8 dynamic response inrush current inrush current is the input current that occurs when the input voltage is first applied. for ac input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied ac voltage. this peak value decreases exponentially during subsequent cycles of ac voltage. overvoltage protection the power supplys overvoltage circuit will be activated when its internal feedback circuit fails. the output voltage shall not exceed its specifications defined on page 3 under protections. power supply will latch off , and require removal/re - application of input ac voltage in order to restart. sho rt circuit protection the power supplys output olp/ocp function also provides protection against short circuits. when a short circuit is applied, the output current will operate in hiccup mode, as shown in the illustration in the olp/ocp section on this page. the power supply will return to normal operation after the short circuit is removed. overload & over current protections the power supplys overload (olp) and over current (ocp) protections will be activated when output current is between 110% and 130% of i o (max load). upon such an occurrence, v o will start to drop. once the power supply has reached its maximum power limit, the protection will be activated. and the power supply will go into hiccup mode (auto - recovery). the power supply will reco ver once the fault condition causing the olp and ocp is removed and i o is back within the specified limit. additionally, if the i o is <130% but >110% for a prolong period of time (depending on the load), the over temperature protection (otp) will be activated due to high temperature on critical components. the power supply will then go into hiccup mode until the fault is removed; and, the input voltage is removed, then reapplied. over temperature protection as mentioned abov e, the power supply also has over temperature protection (otp). this is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. in the event of a higher operating co ndition at 100% load, the power supply will run into otp when the surrounding air temperature is higher than the operating temperature. when activated, the output voltage will go into hiccup mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature. load > 130 % (output hiccup ) 130 % load 0% load t (sec) output voltage (vo) 50% loa d t (sec) 100% load 50ms 50ms 50ms 50ms the power supply output voltage will remain within 5 % of its steady state value, when subjected to a dynamic load change from 50 to 100% of its rated current.
technical d at asheet md s medical ac - dc adapter 24 volt, 1 5 0 watt / mds - 150aas 24 f/b ``` all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 3 ) 9 certificate . all delta medical power products conform to the european directive 2011/65/eu. rohs is the abbreviation for restricti on of the use of certain hazardous substances . delta has been certified as meeting the requirement of iso 13485: 2003 and en iso 13485:2012 for the design and manufacture of switching power supply a nd adaptor for medical device. delta is approved for the ul total certification program (tcp) approved client laboratory for iec60950 and iec60065 . delta also has participated ul client te st data program (c td p) for iec 6060 1. meet doe l evel vi and erp lot 7 efficiency requirement


▲Up To Search▲   

 
Price & Availability of 150AAS24-F

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X